Prolab LIS Cloud

Bidirectional workflow between Prolab LIS and the Beckman Coulter DxI 9000

A Prolab LIS order goes to the Beckman Coulter DxI 9000. The middleware sends tests whose status is Pending (En attente). The same path is used when the laboratory reruns an assay.

Table 1. Statuses that trigger a send.

MomentStatus in ProlabWhat happens next
Order savedPending (En attente)The middleware sends the tests
Send completeActive (Actif)The sample is received on the analyzer
RerunRerun (Refaire) or PendingThe order is sent to the analyzer again

Create the order

From the dashboard

On the dashboard, click New order (Nouvelle demande).

Prolab LIS new laboratory order form, with barcode and Pending status.
Figure 1. New laboratory order, opened from the dashboard.

Select the tests to run, then save. For this analyzer, the group is Immunodosage – Beckman Coulter DXI 9000.

Immunoassay group Beckman Coulter DXI 9000 selected on a Prolab LIS order.
Figure 2. Tests in the immunoassay group, Beckman Coulter DXI 9000.

After you save, the list of tests to send is available. The panel on the right opens ordered tests and results.

Right-hand panel of a Prolab LIS order: ordered tests, results and labels.
Figure 3. Right-hand panel: ordered tests, results and labels.

Send the tests

From Pending to Active

On Ordered tests (Tests demandés), rows in Pending go to the analyzer. The middleware sends them. The technologist does not send them by hand.

Prolab LIS ordered-tests list, assays still in Pending status.
Figure 4. Ordered tests in Pending, before the send.

Once the tests are sent, the status changes to Active (Actif).

Ordered-tests list after the send, with Active status on the assays.
Figure 5. After the send, the status is Active.

On the analyzer, the sample arrives on the Samples screen (Échantillons).

Beckman Coulter DxI 9000 sample list, requested assays shown as ordered.
Figure 6. Sample received on the DxI 9000. The requested assays are listed.

Rerun an assay

Two ways

The technologist can rerun tests in two ways. The analyzer receives the order and runs it again.

Table 2. The two reruns.

Starting pointAction
Ordered testsSelect one or more tests. Change the status from Active to Rerun (Refaire), or to Pending (En attente).
ResultsOpen the assay results. Set the status to Rerun (Refaire).

From the Ordered tests tab, a multi-select changes several rows at once.

Change status menu on a multi-select of ordered tests in Prolab LIS.
Figure 7. Multi-select, then Change status: Rerun or Pending.

From the results, the change is made on the test concerned.

Prolab LIS test results screen with Rerun status on two immunoassay assays.
Figure 8. Status set to Rerun from the test results.

The ordered-test status updates on its own. The middleware then sends the orders to the analyzer again.

Ordered-tests list in Rerun status, ready to be sent again to the analyzer.
Figure 9. Ordered tests in Rerun, ready to be sent again.

The ASTM link and the middleware ports are described in Interfacing Beckman Coulter DxI 9000. All guides are in the documentation.